Chapter 16 SCPI Command Reference
Calculate commands
9102 Handheld Spectrum Analyzer Software version 5.31
325
:CALCulate:MARKer:MREFlevel
:CALCulate:MARKer:FCOunt?
:CALCulate:MARKer:FCOunt:RESolution
:CALCulate:LIMit[:STATe]
Command
The center frequency is changed to the current marker frequency.
Query
There is no query form of this command available.
Example
:CALCulate:MARKer:MCENter.
Syntax
:CALCulate:MARKer:MREFlevel
Parameters
There are no parameters.
Command
The REFerence level is changed to the level at the marker position.
Query
There is no query form of this command available.
Example
:CALCulate:MARKer:MREFlevel.
Syntax
:CALCulate:MARKer:FCOunt?
Parameters
There are no parameters.
Command
There is solely a query form of this command available.
Query
Returns a string with the floating point value for the current result of the fre-
quency counter, in Hz. Marker M1 (CALC:MARK:A COUNter) has to be set
before to get a valid result.
Example
:CALCulate:MARKer:FCOunt?
The value returned in this example is: "2694365000.0"
.
Syntax
:CALCulate:MARKer:FCOunt:RESolution <real1>
Parameters
real1 is an integer. The minimum value for <real1>
is 1, the maximum is
1000
. Valid values are 1, 10, 100, 1000. The default value is 1000.
Command
Sets the counter resolution of the frequency counter in Hz.
Query
Reads and returns the current setting.
Example
:CALC:MARK:FCO:RES 10
:CALC:MARK:FCO:RES?
The value returned in this example is: "10".
Syntax
:CALCulate::LIMit[:STATe] <PredefExpr>
Parameters
PredefExpr is one of the following predefined expressions:
OFF|UPPer|LOWer|UPPLow.
Default is OFF.