Chapter 16 SCPI Command Reference
Calculate commands
9102 Handheld Spectrum Analyzer Software version 5.31
321
:DISPlay:COLor:GRATicule
:DISPlay:COLor:LIMits
Calculate commands
The markers of the 9102 can be set using the calculate commands.
:CALCulate:MARKer:AOFF
:CALCulate:MARKer:{A|B|C|D|E|F}[:STATe]
Syntax
:DISPlay:COLor:GRATicule <int1>
Parameters
int1 is an integer. The minimum value for <int1> is 1, the maximum is 8.
The default value is 2.
Command
Sets the color from the color palette for the graticule.
Query
Reads and returns the current setting.
Example
:DISPlay:COLor:GRATicule 5
:DISPlay:COLor:GRATicule?
The value returned in this example is: "5".
Syntax
:DISPlay:COLor:LIMits <int1>
Parameters
int1 is an integer. The minimum value for <int1> is 1, the maximum is 8.
The default value is 7.
Command
Sets the color from the color palette for the limit lines.
Query
Reads and returns the current setting.
Example
:DISPlay:COLor:LIM 5
:DISP:COL:LIMits?
The value returned in this example is: "5".
Syntax
:CALCulate:MARKer:AOFF
Parameters
There are no parameters.
Command
All markers are switched off.
Query
There is no query form of this command available.
Example
:CALC:MARK:A:STAT NORM
:CALC:MARK:B:STAT DELT
:CALC:MARK:C:STAT DELT
:CALC:MARK:AOFF
Syntax
:CALCulate:MARKer:{A|B|C|D|E|F}[:STATe]
<PredefExpr>
Parameters
PredefExpr is one of the following predefined expressions:
OFF|NORMal|DELTa.
Default is OFF.