Chapter 16 SCPI Command Reference
MMemory commands
9102 Handheld Spectrum Analyzer Software version 5.31
299
:INPut:EDEVice
:INPut:AFACtor
:INPut:CFACtor
MMemory commands
With the MMemory commands, you can fully exploit the capabilities of
the instrument to store and reload measurement results in its nonvola-
tile memory.
:MMEMory:STORe:STATe
Syntax
:INPut:EDEVice <PredefExpr>
Parameters
PredefExpr is one of the following predefined expressions: ON|OFF.
Default is OFF.
Command
Switches the external device compensation on or off. A corresponding file
has to be loaded first with command MMEMory:LOAD:EDEVice.
Query
Reads and returns the current setting.
Example
:INP:EDEVice ON
:INPut:EDEVice?
Returns the following string: "ON"
Syntax
:INPut:AFACtor <PredefExpr>
Parameters
PredefExpr
is one of the following predefined expressions: OFF|ON.
Default is OFF.
Command
Switches the antenna factor compensation on or off. A corresponding file
has to be loaded first with command MMEMory:LOAD:AFACtor.
Query
Reads and returns the current setting.
Example
:INPut:AFACtor ON
:INPut:AFACtor?
Returns the following string:
"ON"
Syntax
:INPut:CFACtor <PredefExpr>
Parameters
PredefExpr
is one of the following predefined expressions: OFF|ON.
Default is OFF.
Command
Switches the cable factor compensation on or off. A corresponding file has
to be loaded first with command MMEMory:LOAD:CFACtor.
Query
Reads and returns the current setting.
Example
:INPut:CFACtor ON
:INPut:CFACtor?
Returns the following string: "ON"
Syntax
:MMEMory:STORe:STATe <string1>
Parameters
string1 is a string (text) parameter. The maximum length of string1 is
11 characters.