Chapter 16 SCPI Command Reference
Calculate commands
324 9102 Handheld Spectrum Analyzer Software version 5.31
:CALCulate:{A|B|C|D|E|F}:MARKer:FSTep
:CALCulate:MARKer:{A|B|C|D|E|F}:TSELect
:CALCulate:MARKer:MAXPeak
:CALCulate:MARKer:NPEak
:CALCulate:MARKer:MCENter
Syntax
:CALCulate:MARKer:{A|B|C|D|E|F}:FSTep
Parameters
There are no parameters.
Command
Sets Marker A, B, C, D, E or F to current frequency step (FSTep).
Query
There is no query form of this command available.
Example
:CALCulate:MARKer:A:FSTep
Syntax
:CALCulate:MARKer:{A|B|C|D|E|F}:TSELect <PredefExpr>
Parameters
PredefExpr is one of the following predefined expressions: A|B
.
Default is A.
Command
Selects an active marker and sets it either to trace A or trace B. Note: Set to
trace B is only possible when trace B is active.
Query
The query form of this command will return the current setting. The string
delivered back will contain the short-form version of one of the predefined
expressions explained above.
Example
:CALC:MARK:A:TSELect B
:CALC:MARK:A:TSELect?
Value returned in this example: "B"
.
Syntax
:CALCulate:MARKer:MAXPeak
Parameters
There are no parameters.
Command
Sets the currently selected marker to the maximum measured level. A
marker is "selected" by way of the :CALCu-
late:MARKer:{A|B|C|D|E|F}[:STATe] command.
Query
There is no query form of this command available.
Example
:CALCulate:MARKer:MAXPeak.
Syntax
:CALCulate:MARKer:NPEak
Parameters
There are no parameters.
Command
Sets the currently selected marker to the next highest level value.
Query
There is no query form of this command available.
Example
:CALCulate:MARKer:NPEak.
Syntax
:CALCulate:MARKer:MCENter
Parameters
There are no parameters.