Chapter 13 Troubleshooting
Recharging problems with the 9102B
214 9102 Handheld Spectrum Analyzer Software version 5.31
Recharging problems with the 9102B
If a 9102B (serial number 0604001 and higher) is completely
discharged then it may happen that the 9102B does not automatically
re-charge the battery while connected to an external power supply.
In such a case, to reset the "safety power-off switch", please carry out
the following steps:
1Switch off the 9102B.
2 Remove the power connector from the 9102B.
3 Remove the battery.
4 Wait a few seconds.
5 Insert the battery again.
6 Re-connect the power connector and provide external power.
The LED on the front will show that the battery is charged again.