Chapter 16 SCPI Command Reference
Overview
258 9102 Handheld Spectrum Analyzer Software version 5.31
Overview
The command set of the 9102 Handheld Spectrum Analyzer follows
the SCPI standard and is broken down into the following subsystems:
General commands
System commands
Sense commands
Input commands
Memory commands
Display commands
Calculate commands
Format commands
Each keyword in the command can be used either in its long or its short
form. Uppercase letters are used to indicate the short-form command
syntax. Within one keyword, either the short form or the full version can
be used, but not a mix.
The syntax can be used to form either a command or a query (with a
question mark behind the command form). Many SCPI commands have
either a command form or a query form, but not both.
Please refer to the “Index of SCPI Commands” on page 351 for an
alphabetical list of commands.
General commands
:REBoot
:HCOPy[:IMMediate]
Syntax
:REBoot
Parameters
There are no parameters.
Command
Reboots the 9102. The current settings are not affected by this command.
Query
There is no query form of this command available.
Example
:REBoot
Syntax
:HCOPy[:IMMediate]
Parameters
There are no parameters.
Command
Starts a printout over the serial port. For further informations see command
:SYSTem:PRINter:TYPe
.
Query
There is no query form of this command available.
Example
:HCOPy