Chapter 16 SCPI Command Reference
Calculate commands
9102 Handheld Spectrum Analyzer Software version 5.31
329
:CALCulate:MEASure:EMF
:CALCulate:MEASure:PSENsor[:ALL]?
:CALCulate:MEASure:PSENsor:FPOWer?
Parameters
There are no parameters.
Command
There is solely a query form of this command available.
Query
Returns the average cable loss, in dB. This query returns valid results only
in Cable Loss mode.
Example
:CALCulate:MEASure:ACLoss?
The value returned in this example is: "-22.5"
.
Syntax
:CALCulate:MEASure:EMF?
Parameters
There are no parameters.
Command
There is solely a query form of this command available
Query
Returns the EMF measurement value. The result depends on the setting of
the :SENSe:EMF:MEASure:DISPlay command. If it is set to
EFSTrength
(electric field strength) the unit for the result is Volt/meter. If it
is set to PDENsity
(power density) the unit for the result is Watt/squa-
remeter. This query only returns valid results in EMF mode.
Example
:CALCulate:MEASure:EMF?
The value returned in this example is: "0.103352"
.
Syntax
:CALCulate:MEASure:PSENsor[:ALL]?
Parameters
There are no parameters.
Command
There is solely a query form of this command available.
Query
Returns the power sensor measurement values. The first result represent
the forward power, the second result represent the reverse power. The
results depends on the setting :SENSe:PSENSor:UNIT command. If it is
set to DBM the unit for the result is dBm. If it is set to W the unit for the result
is Watt. This query returns valid results only in power sensor mode.
Example
:CALCulate:MEASure:PSENsor?
The value returned in this example is: "20.1,18.5".
Syntax
:CALCulate:MEASure:PSENsor:FPOWer?
Parameters
There are no parameters.
Command
There is solely a query form of this command available.
Query
Returns the power sensor measurement value for the forward power. The
result depends on the setting :SENSe:PSENSor:UNIT command. If it is
set to DBM the unit for the result is dBm. If it is set to W the unit for the result
is Watt. This query returns valid results only in power sensor mode.
Example
:CALCulate:MEASure:PSENsor:FPOWer?
The value returned in this example is: "20.1".