EasyManua.ls Logo

Aeroflex 9102 - Chapter 14 Updating the Instrument Software

Aeroflex 9102
416 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 12 EMF (EMI) Operation
EMF measurements with the 9102
9102 Handheld Spectrum Analyzer Software version 5.31
195
measurements with the exception of connecting the antennas to the
9102. The steps involved in this procedure depend on the antenna
used.
NOTE
Due to the physical nature of EMF measurements the following has
to be taken into account.
When performing measurements on broadband signals the
electrical field strength or the power density displayed may show
deviations. In this case the measurement has to be performed via a
channel power measurement. Furthermore a very big or very small
relation between span and RBW may lead to false measurements or
invalid measurement results.

Table of Contents