Chapter 16 SCPI Command Reference
General commands
9102 Handheld Spectrum Analyzer Software version 5.31
263
*STB
Example
*SRE 68
As soon as an error occurs, bits 2 and 6 of the service register will be set.
ANDed with the mask (68), a binary 1 will be the result.
Syntax
*STB?
Parameters
There are no parameters.
Command
There is only a query form of this command available.
Query
Returns the decimal representation of the current contents of the service
register in a string.
Note: This register is self-destructive, i.e. its contents will be cleared after
reading.
Example
A *STB? command returns "68".
The return value of 68 (= 64 + 4) means that an error occurred (4).