Chapter 16 SCPI Command Reference
Sense commands
284 9102 Handheld Spectrum Analyzer Software version 5.31
:SENSe:REFLevel:OFFSet
:SENSe:STATe
:SENSe:MEASure
:SENSe:MEASure:OBW
Example
:SENSe:REFLevel:UNIT DBMV
:SENSe:REFLevel:UNIT?
Returns the following string: "DBMV"
Syntax
:SENSe:REFLevel:OFFSet <PredefExpr>
Parameters
real1 is a floating point real number.
The minimum value for real1 is -113.0, the maximum value 9999999.9.
real1 can be set in multiples of 1000. The default value for real1 is
5000000.
Command
Defines the reference level offset.
Query
Returns the current setting.
Example
:SENSe:REFLevel:OFFset -10
:SENSe:REFLevel:OFFSet?
Returns the following string: "-10"
Syntax
:SENSe:STATe?
Parameters
There are no parameters.
Command
There is solely a query form of this command available.
Query
Reads and returns the Uncal state.
Example
:SENSe:STATe?
Returns the following string:
"ON".
Syntax
:SENSe:MEASure <PredefExpr>
Parameters
PredefExpr is one of the following predefined expressions:
NONe|CPOWer|ACPR|OBW. Default is NONe.
Command
Starts measurements of the type described by PredefExp.
Query
Reads and returns the current setting.
Example
:SENSe:MEASure ACPR
:SENSe:MEASure?
Returns the following string: "ACPR"
Syntax
:SENSe:MEASure:OBW <int1>
Parameters
int1 is an integer. The minimum value for int1 is 5, the maximum is 99.
The default value for int1 is 90.
Command
Sets the percentage value for which the 9102 shall determine the occupied
bandwidth (spectrum analyzer mode).
Query
Reads and returns the current setting.