15 UART/RS232 Triggering and Serial Decode
336 InfiniiVision 7000B Series Oscilloscopes User’s Guide
Note that the setting of this softkey does not affect the selected base of
the decode display.
3 Press the Trigger Setup softkey and set up the desired trigger condition:
Rx Start Bit — The oscilloscope triggers when a start bit occurs on Rx.
Rx Stop Bit — Triggers when a stop bit occurs on Rx. The trigger will
occur on the first stop bit. This is done automatically whether the
device under test uses 1, 1.5, or 2 stop bits. You do not need to
specify the number of stop bits used by the device Under test.
Rx Data — Triggers on a data byte that you specify. For use when the
device under test data words are from 5 to 8 bits in length (no 9th
(alert) bit).
Rx 1:Data — For use when the device under test data words are 9 bits
in length including the alert bit (the 9th bit). Triggers only when the
9th (alert) bit is 1. The specified data byte applies to the least
significant 8 bits (excludes the 9th (alert) bit).
Rx 0:Data — For use when the device under test data words are 9 bits
in length including the alert bit (the 9th bit). Triggers only when the
9th (alert) bit is 0. The specified data byte applies to the least
significant 8 bits (excludes the 9th (alert) bit).
Rx X:Data — For use when the device under test data words are 9 bits
in length including the alert bit (the 9th bit). Triggers on a data byte
that you specify regardless of the value of the 9th (alert) bit. The
specified data byte applies to the least significant 8 bits (excludes
the 9th (alert) bit).
Similar choices are available for Tx.
Rx or Tx Parity Error — Triggers on a parity error based on the parity
you have set in the Bus Configuration Menu.
4 If you choose a trigger condition that includes “Data” in its description
(for example: Rx Data), then press the Data is softkey, and choose an
equality qualifier. You can choose equal to, not equal to, less than, or
greater than a specific data value.
5 Use the Data softkey to choose the data value for your trigger
comparison. This works in conjunction with the Data is softkey.
6 Optional: The Burst softkey lets you trigger on the Nth frame (1- 4096)
after an idle time you select. All trigger conditions must be met for the
trigger to occur.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7000B Series and is the answer not in the manual?

Agilent Technologies 7000B Series Specifications

General IconGeneral
BrandAgilent Technologies
Model7000B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals