Optimizing B-Mode
LOGIQ 7 Online Help 5-13
Direction 2392536-100 Rev. 1
CrossBeam (continued)
Values All linear and curved linear (for future validation) probes are
supported. Multiple focal zones are supported. CrossBeam is
available while in B-Mode. Steering is optimized by probe. The
displayed compound image depth is equal to the image depth of
the non-steered frame.
The following features are supported in CrossBeam:
• Harmonic Imaging
• CINE Loops acquired using CrossBeam.
• Read Zoom
• All Measurement and Analysis packages
• Line Density, Depth, Map, TGC, Gain, Edge Enhance,
Frame Averaging, Compression, Colorize, Acoustic Output,
Rejection, Rotation, B-Softener, Auto, Reverse, Dynamic
Range, Biopsy
The following controls are not available:
• Anatomical M-Mode
• LOGIQ View
• 3D Imaging
• B Flow
Benefits The combined single image has the benefits of reduced speckle
noise, reduced clutter, and continuity of specular reflectors.
Therefore, this technique can improve contrast resolution with
increased conspicuity of low contrast lesions, better detection of
calcifications, biopsy needle visualization, and cystic boundary
definition.