System Care and Maintenance
LOGIQ 7 Online Help 18-21
Direction 2392536-100 Rev. 1
Other Maintenance
Replacing illuminated key caps/lamps
Contact a local Service Representative when a key cap or lamp
needs to be replaced.
Uninterruptible Power Supply (UPS) Option
Lifetime The lifetime of a UPS is limited and varies according to
circumstances. It lasts for three years under normal conditions;
however, it may last one year if the UPS is used over twelve
hours per day and if the room temperature is over 30 degrees C.
Possible damage may occur with the HDD, system software and
user data by the electrical power outage or by disconnecting the
power plug after the battery is dead.
Battery
Replacement and
Disposition
Battery replacement every three years is recommended.
Contact a local Service Representative for the replacement of
the battery. Used batteries will be discarded appropriately by
GE.
NOTE: Disposing of the battery should meet local law and regulatory
requirements.
NOTE: Dispose of the system according to local law and regulatory
requirements.