EasyManuals Logo

GE LOGIQ 7 Basic User Manual

GE LOGIQ 7
984 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #240 background imageLoading...
Page #240 background image
Optimizing the Image
5-100 LOGIQ 7 Online Help
Direction 2392536-100 Rev. 1
Angle Correct
Description Estimates the flow velocity in a direction at an angle to the
Doppler vector by computing the angle between the Doppler
vector and the flow to be measured.
NOTE: When the Doppler Mode Cursor and angle correct indicator are
aligned (the angle is O), you cannot see the angle correct
indicator.
Adjusting Flow toward the probe is mapped above the baseline and right
speaker (as you are facing the system).
Flow toward the probe is mapped above the baseline.
To adjust the angle relative to the probe face, adjust Angle
Correct to the left/right. The velocity scale changes when you
adjust angle correct.
Values 1 increments from 0 to 90. The possible range of operation is
from 0 degrees to 90 degrees in either direction. For optimum
velocity measurements, the angle of incidence should be
between 45-60 for vascular applications.
Angle Correct values vary by probe and application. Angle
Correct values are returned to the factory or user preset value
when you change the following: Probe, Exam Category, Exam
Calcs, or New Patient.
Benefits Optimizes the accuracy of the flow velocity. This is especially
useful in vascular applications where you need to measure
velocity.
Quick Angle
Description Quickly adjusts the angle by 60 degrees.
Adjusting To quickly adjust the angle, press Quick Angle.
Values 0 degrees and 60 degrees.

Table of Contents

Other manuals for GE LOGIQ 7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ 7 and is the answer not in the manual?

GE LOGIQ 7 Specifications

General IconGeneral
BrandGE
ModelLOGIQ 7
CategoryMedical Equipment
LanguageEnglish

Related product manuals