Preparing the System for Use
3-4 LOGIQ 7 Basic User Manual
Direction 2392206-100 Rev. 1
Environmental Requirements
The system should be operated, stored, or transported within
the parameters outlined below. Either its operational
environment must be constantly maintained or the unit must be
turned off.
Table 3-1: System Environmental Requirements
Operational Storage Transport (<16hrs.)
Temperature 10° - 35°C
50° - 95°F
-10° - 60°C
14° - 140°F
-40° - 60°C
-40° - 140°F
Humidity 30 - 80% non-condensing 30 - 90% non-condensing 30 - 90%
non-condensing
Pressure 700 - 1060hPa 700 - 1060hPa 700 - 1060hPa
WARNING
To ensure the UPS battery remains fully charged, DO NOT turn
off the breaker switch or disconnect the system power cord
from the AC outlet, except for maintenance or during a portable
exam.