Probes and Biopsy
17-10 LOGIQ 7 Online Help
Direction 2392536-100 Rev. 1
Probe Usage
For details on connecting, activating, deactivating,
disconnecting, transporting and storing the probes,
Care and Maintenance
Inspecting probes
Perform After
Each Use
Inspect the probe's lens, cable, and casing. Look for any
damage that would allow liquid to enter the probe. If any
damage is found, do not use the probe until it has been
inspected and repaired/replaced by a GE Service
Representative.
NOTE: Keep a log of all probe maintenance, along with a picture of any
probe malfunction.
Environmental Requirements
Probes should be operated, stored, or transported within the
parameters outlined below.
CAUTION
Ensure that the probe face temperature does not exceed the
normal operation temperature range.
Table 17-5: Probe Environmental Requirements
Operational Storage Transport
Temperature 10° - 40° C
50° - 104° F
-10° - 60° C
14° - 140° F
-40° - 60° C
-40° - 140° F
Humidity 30 - 85 %
non-condensing
30 - 90%
non-condensing
30 - 90%
non-condensing
Pressure 700 - 1060hPa 700 - 1060hPa 700 - 1060hPa