Optimizing the Image
5-106 LOGIQ 7 Online Help
Direction 2392536-100 Rev. 1
Trace Direction
Description Specifies trace direction.
Adjusting Press Trace Direction.
Values Above, Below, or Both.
Benefits You can select where on the waveform to perform the trace,
above, below, or both.
Full Timeline
Description Expands display to full timeline display.
Adjusting Press Full Timeline.
Values On/Off.
Benefits Allows you to view just the timeline, to see more detail.
Display Format
Description Changes the horizontal/vertical layout between B-Mode and M-
Mode, or timeline only.
Adjusting Press Display Format.
Values Vertical 1/3, 1/2, or 2/3 B-Mode, Horizontal 1/4 or 1/2 B-Mode, or
Timeline Only.
Benefits You can select how to have your Doppler timeline and anatomy
displayed.
Time Resolution
Description Adjusts image appearance so that if you select a lower setting,
the image appears smoother; if you select a higher setting, the
image appears sharper.
Adjusting Press Time Resolution.
Values 1-3
Benefits Time Resolution = 3 (high) smooths the Timeline; Time
Resolution =1 (low) increases the time resolution (image not as
smooth).

Table of Contents

Other manuals for GE LOGIQ 7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ 7 and is the answer not in the manual?

GE LOGIQ 7 Specifications

General IconGeneral
BrandGE
ModelLOGIQ 7
CategoryMedical Equipment
LanguageEnglish

Related product manuals