Optimizing the Image
5-12 LOGIQ 7 Online Help
Direction 2392536-100 Rev. 1
CrossBeam
Description CrossBeam combines multiple co-planar images from different
view angles into a single image at real-time frame rates, using
bi-cubic interpolation.
Adjusting CrossBeam combines frames steered from a minimum of three
different angles and up to a maximum of five steer angles.
To activate CrossBeam, press the CrossBeam key on the
Touch Panel.
To adjust the number of frames being compounded, press the
CrossBeam# Touch Panel key. Select Low or High.
You can preset CrossBeam to activate upon entering B-Mode;
you can also preset the CrossBeam # via the Utility, Imaging, B-
Mode page.