Preparing for a Biopsy
LOGIQ 7 Online Help 17-41
Direction 2392536-100 Rev. 1
Post Biopsy
When the biopsy is complete, remove the needle barrel, needle
clip and probe sheath. Properly dispose of these items in
accordance with current facility guidelines.
The biopsy bracket can be sterilized in a recommended
disinfecting agent and reused.
E8C Probe Biopsy Guide
When the E8C probe is attached and active, the needle guide
type is the TR5° (Civco disposable guide with a 5° offset angle).
Figure 17-24. TR5° Biopsy Guide
Preparation
To prepare the E8C for use:
1. Remove the probe from the box and carefully examine it for
any damage.
2. If the biopsy guide is to be attached, use the filling removal
tool to clean out the attachment area on the probe head.
Figure 17-25. Attachment Filling Removal
a. Probe Head
b. Attachment
c. Filling Removal Tool
3. Clean, then disinfect the probe.
NOTE: Ensure that protective gloves are worn.
b