Cardiology
10-28 LOGIQ 7 Basic User Manual
Direction 2392206-100 Rev. 1
M-Mode Measurements
Aorta
The following are cardiac M-Mode aorta measurements:
One distance
measurement
• Aortic Root Diameter (Ao Root Diam)
Making a one
distance
measurement
1. Select the measurement; an active caliper displays.
2. To position the active caliper at the start point, move the
Trackball.
3. To fix the start point, press Set.
The system fixes the first caliper and displays a second
active caliper.
4. To position the second active caliper at the end point, move
the Trackball.
A dotted line connects the measurement points.
5. To complete the measurement, press Set.
The system displays the measurement in the Results
Window.