Preparing for a Biopsy
LOGIQ 7 Online Help 17-39
Direction 2392536-100 Rev. 1
Biopsy Need Path Verification
To verify that the path of the needle is accurately indicated within
the guidezone on the system monitor, perform the following:
• Properly install the bracket and biopsy guide.
• Scan in a container filled with water (47° C).
• Display the biopsy guidezone on the monitor.
• Ensure that the needle echo falls within the guidezone
markers.
To verify that the path of the needle is accurately indicated within
the guidezone on the system monitor, perform the following:
• Properly install the bracket and biopsy guide.
• Scan in a container filled with water (47° C).
• Display the biopsy guidezone on the monitor.
• Ensure that the needle echo falls within the guidezone
markers.