Optimizing the Image
5-68 LOGIQ 7 Online Help
Direction 2392536-100 Rev. 1
Optimizing M Color Flow
M Color Flow Mode
Description M Color Flow is used for fetal cardiac applications. Color Flow
overlays color on the M-Mode image using velocity and variance
color maps. The Color Flow wedge overlays the B-Mode image
and M-Mode timeline.
The Color Flow maps available in M-Mode are the same as in
Color Flow Mode. The size and position of the Color Flow
window in B-Mode determines the size and position of the Color
Flow window in M-Mode.
All M-Mode measurements are available with M Color Flow
active: depth, distance along a straight line, % stenosis, volume,
trace, circumference, enclosed area, distance, time, slope, and
heart rate.
Activating To activate M Color Flow Mode, press M (M-Mode). Then press
CF (Color Flow) - or - press CF, then press M.
To toggle between M Color Flow controls and Color Flow
controls, press the appropriate Touch Panel Mode tab.
Values On/Off.
Benefits Color Flow Mode and Color M-Mode are Doppler Modes
intended to add color-coded qualitative information concerning
the relative velocity and direction of fluid motion within the B-
Mode or M-Mode image.