Safety
2-16 LOGIQ 7 Basic User Manual
Direction 2392206-100 Rev. 1
EMC Performance (continued)
Portable and mobile radio communications equipment (e.g. two-
way radio, cellular/cordless telephones, wireless computer
networks), other than those supplied by GE (wireless
microphone for VoiceScan), should be used no closer to any
part of this system, including cables, than determined according
to the following method:
CAUTION
Do not use devices which intentionally transmit RF signals
(cellular phones, transceivers, or radio controlled products),
other than those supplied by GE (wireless microphone, for
example) unless intended for use with this system, in the
vicinity of this equipment as it may cause performance outside
the published specifications.
Keep power to these devices turned off when near this
equipment.
Medical staff in charge of this equipment is required to instruct
technicians, patients and other people who may be around this
equipment to fully comply with the above regulation.
Table 2-5: Portable and mobile radio communications equipment distance requirements
Frequency Range: 150 kHz - 80 MHz 80 MHz - 800 MHz 800 MHz - 2.5 GHz
Calculation Method: d=[3.5/V
1
] square root
of P
d = [3.5/E
1
] square root
of P
d = [7/E
1
] square root of
P
Where: d= separation distance in meters, P = rated power of the transmitter, V
1
=compliance value for
conducted RF, E
1
= compliance value for radiated RF
If the maximum
transmitter power in
watts is rated
The separation distance in meters should be
5 2.6 2.6 5.2
20 5.2 5.2 10.5
100 12.0 12.0 24.0