OB/GYN
9-38 LOGIQ 7 Basic User Manual
Direction 2392206-100 Rev. 1
Peak Systole (PS), End Diastole (ED), or Minimum Diastole (MD)
To calculate the peak systole, end diastole, or minimum diastole:
1. Select PS, ED, or MD.
The system displays an active caliper with vertical and
horizontal dotted lines.
2. To position the caliper at the measurement point, move the
Trackball.
3. To complete the measurement, press Set.
The system displays the peak systole, end diastole, or
minimum diastole in the Results Window.
PS/ED or ED/PS Ratio
To calculate the Peak Systole/End Diastole ratio or End
Diastole/Peak Systole ratio:
1. Select PS/ED or ED/PS.
The system displays an active caliper with vertical and
horizontal dotted lines.
2. To position the caliper at peak systole (PS) or end diastole
(ED), move the Trackball.
3. To fix the measure point, press Set.
The system displays a second active caliper.
4. To position the second caliper at end diastole (ED) or peak
systole (PS), move the Trackball.
5. To complete the measurement, press Set.
The system displays the peak systole, end diastole, and PS/
ED or ED/PS ratio in the Results Window.

Table of Contents

Other manuals for GE LOGIQ 7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ 7 and is the answer not in the manual?

GE LOGIQ 7 Specifications

General IconGeneral
BrandGE
ModelLOGIQ 7
CategoryMedical Equipment
LanguageEnglish

Related product manuals