Safety
2-20 LOGIQ 7 Basic User Manual
Direction 2392206-100 Rev. 1
Peripheral Update for EC countries (continued)
General precautions for installing an alternate on-board device
would include:
1. The added device must have appropriate safety standard
conformance and CE Marking.
2. The total power consumption of the added devices, which
connect to the LOGIQ 7 and are used simultaneously, must
be less than or equal to the rated supply of the LOGIQ 7.
3. There must be adequate heat dissipation and ventilation to
prevent overheating of the device.
4. There must be adequate mechanical mounting of the device
and stability of the combination.
5. Risk and leakage current of the combination must comply
with IEC/EN 60601-1.
6. Electromagnetic emissions and immunity of the combination
must conform to IEC/EN 60601-1-2.
General precautions for installing an alternate off-board, remote
device or a network would include:
1. The added device(s) must have appropriate safety standard
conformance and CE Marking.
2. The added device(s) must be used for their intended
purpose having a compatible interface.
3. Signal or mains isolation devices and additional protective
earth may be needed to assure compliance with IEC/
EN 60601-1-1.
CAUTION
The connection of equipment or transmission networks other
than as specified in the user instructions can result in an
electric shock hazard or equipment malfunction. Substitute or
alternate equipment and connections requires verification of
compatibility and conformity to IEC/EN 60601-1-1 by the
installer. Equipment modifications and possible resulting
malfunctions and electromagnetic interference are the
responsibility of the owner.