Cardiology
10-10 LOGIQ 7 Basic User Manual
Direction 2392206-100 Rev. 1
Aortic Valve
The following are cardiac B-Mode aortic valve measurements:
One distance
measurements
• Aortic Diameter (ASD Diam)
• Aortic Valve Cusp Separation (AV Cusp)
• Aortic Valve Diameter (AV Diam)
Making a one
distance
measurement
1. Select the measurement; an active caliper displays.
2. To position the active caliper at the start point, move the
Trackball.
3. To fix the start point, press Set.
The system fixes the first caliper and displays a second
active caliper.
4. To position the second active caliper at the end point, move
the Trackball.
A dotted line connects the measurement points.
5. To complete the measurement, press Set.
The system displays the measurement in the Results
Window.
One trace
measurements
• Aortic Valve Area Planimetry (AVA Planimetry)
• Transverse Aortic Valve Area (Trans AVA(d), Trans
AVA(s))
Making a trace
measurement
1. Select the measurement; an active caliper displays.
2. To position the caliper at the start point, move the Trackball.
3. To fix the trace start point, press Set.
The system displays a vertical dotted line.
4. To trace the measurement area, move the Trackball.
A line shows the traced area.
5. To complete the measurement, press Set.
The system displays the measurement in the Results
Window.

Table of Contents

Other manuals for GE LOGIQ 7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ 7 and is the answer not in the manual?

GE LOGIQ 7 Specifications

General IconGeneral
BrandGE
ModelLOGIQ 7
CategoryMedical Equipment
LanguageEnglish

Related product manuals