EasyManuals Logo

GE LOGIQ 7 Basic User Manual

GE LOGIQ 7
984 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #542 background imageLoading...
Page #542 background image
Cardiology
10-26 LOGIQ 7 Basic User Manual
Direction 2392206-100 Rev. 1
System
The following are cardiac B-Mode System measurements:
One distance
measurements
• Interventricular Septum Thickness
• Diastolic (IVSd)
• Systolic (IVSs)
• Inferior Vena Cava (IVC)
• Main Pulmonary Artery Diameter (MPA)
• Systemic Vein Diameter (Systemic Diam)
• Patent Ductus Arterosis Diameter (PDA Diam)
• Patent Foramen Ovale Diameter (PFO Diam)
• Pericard Effusion Diastole (PEd)
• Ventricular Septal Defect Diameter (VSD Diam)
Making a one
distance
measurement
1. Select the measurement; an active caliper displays.
2. To position the active caliper at the start point, move the
Trackball.
3. To fix the start point, press Set.
The system fixes the first caliper and displays a second
active caliper.
4. To position the second active caliper at the end point, move
the Trackball.
A dotted line connects the measurement points.
5. To complete the measurement, press Set.
The system displays the measurement in the Results
Window.
Body surface area
and stroke volume
measurements
• Interventricular Septum (IVS) Fractional Shortening (IVSs)
The system calculates body surface area from the patient’s
height and weight.

Table of Contents

Other manuals for GE LOGIQ 7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ 7 and is the answer not in the manual?

GE LOGIQ 7 Specifications

General IconGeneral
BrandGE
ModelLOGIQ 7
CategoryMedical Equipment
LanguageEnglish

Related product manuals