Cardiology
10-30 LOGIQ 7 Basic User Manual
Direction 2392206-100 Rev. 1
Left Atrium
The following are cardiac M-Mode Left Atrium measurements:
Two distance
measurement
(ratio)
• Left Atrium Diameter to AoRoot Diameter Ratio (LA/Ao)
Making two
distance
measurements
1. Select LA/Ao; an active caliper displays.
2. Make the first distance measurement:
a. To position the active caliper at the start point, move the
Trackball.
b. To fix the start point, press Set.
The system fixes the first caliper and displays a second
active caliper.
c. To position the second active caliper at the end point,
move the Trackball.
A dotted line connects the measurement points.
d. To complete the measurement, press Set.
The system displays the distance value in the Results
Window. After the first measurement, the system
displays an active caliper.
3. To make the second distance measurement, repeat steps
a–d above.
The system displays the measurements and ratio in the
Results Window.