Cardiology
10-38 LOGIQ 7 Basic User Manual
Direction 2392206-100 Rev. 1
Pulmonic Valve
The following are cardiac M-Mode Pulmonic Valve
measurements:
One time interval
measurements
QRS complex to end of envelope (Q-to-PV close)
Making a time
interval
measurement
1. Select the measurement.
The system displays an active caliper and a vertical dotted
line.
2. To position the caliper at the start point, move the Trackball.
3. To fix the first caliper, press Set.
The system displays a second active caliper.
4. To position the caliper at the end point, move the Trackball.
5. To complete the measurement, press Set.
The system displays the time interval in the Results
Window.