ECG Display
Description Provides the ability to turn on the ECG trace and Auto Heart
Rate for display on the monitor.
Adjusting When the key is selected, the ECG trace and Auto Heart Rate
toggle between on and off.
Value On or Off
ECG Trigger Period
Description The control specifies the number of heart cycles (R-waves) that
are skipped between ECG triggers.
The default is 1 or no skipping.
Adjusting Adjust the corresponding control.
Value 1 - 30
Delay Time
Description In ECG Trigger Mode: If only ECG Trig1 or ECG Trig2 is
selected via the ECG Trigger key, the Delay Time key controls
the R-Delay time of the active trigger. If both triggers are
selected, the primary control is ECG Trig1, but adjusting this
control switches to ECG Trig2. Once the trigger is set, the snap
shot image is displayed each time the update line passes the
active trigger(s).
In Timer Trigger Mode: Rotating the knob changes the delay
time between images.
Adjusting Adjust the corresponding control.
Value 0.10 - 10.00