Urology Calculations
LOGIQ 7 Basic User Manual 12-7
Direction 2392206-100 Rev. 1
Renal Volume
This calculation uses a standard distance measurement. Length
is typically measured in the sagittal plane. Width and height are
measured in the axial plane.
Figure 12-4. Renal Volume Touch Panel
To measure Renal Volume:
1. Scan the patient in the appropriate scan plane.
2. Select the Renal folder, an active caliper displays.
3. Perform a standard distance measurement:
The system displays the distance value in the Results
Window.
4. To make the second and third distance measurement,
repeat steps 2–3.
After you complete the third distance measurement, the
system displays the renal volume in the Results Window.