Powering the System
LOGIQ 7 Basic User Manual 3-17
Direction 2392206-100 Rev. 1
Acclimation Time
After being transported, the unit requires one hour for each 2.5°
increment its temperature is below 10° C or above 40° C.
Table 3-3: System Acclimation Time Chart
° C 60 55 50 45 40 35 30 25 20 15 10
° F 140 131 122 113 104 95 86 77 68 59 50
hours 86420000000
° C 5 0 -5 -10 -15 -20 -25 -30 -35 -40
° F 41 32 23 14 5 -4 -13 -22 -31 -40
hours 2468101214161820