Quality Assurance
LOGIQ 7 Online Help 18-33
Direction 2392536-100 Rev. 1
Lateral distance measurements
Description Lateral measurements are distance measurements obtained
perpendicular to the axis of the sound beam. See Figure 18-11
for more information.
Benefit The purpose is the same as vertical measurements. Precisely-
spaced horizontal pin targets are scanned and results compared
to the known distance in the phantom.
Method Lateral distance should be measured in the near, mid and far
fields as well as in zoom. If necessary, different depths of fields
of view can be tested.
Procedure To measure lateral distance:
1. Scan a test phantom with precisely-spaced horizontal pin
targets. Adjust all scan controls, as necessary, for the best
image of the pin targets from side to side.
2. Press Freeze to stop image acquisition and perform a
standard distance measurement between the pins at
different points in the image. Record all images for
archiving.
3. Scan the horizontal pins in zoom or at different depth/scale
factors.
4. Press Freeze to stop image acquisition; repeat the distance
measurements between pins and record the images for
archiving.
5. Document the measurements for reference and future
comparison.
Contact a Service Engineer if horizontal measurements differ by
more than 3mm or 3% of that depth, whichever is greater.

Table of Contents

Other manuals for GE LOGIQ 7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ 7 and is the answer not in the manual?

GE LOGIQ 7 Specifications

General IconGeneral
BrandGE
ModelLOGIQ 7
CategoryMedical Equipment
LanguageEnglish

Related product manuals