Test Sequence Application 7
Keysight U8903B User’s Guide 305
Coupling
Press the Coupling softkey to select the coupling type.
–DC
DC coupling allows both AC and DC input signals to pass through to the Bluetooth analyzer and to be measured down to
0 Hz. This setting should be selected when making DC voltage measurements.
–AC
AC coupling blocks the DC component of the input signal. This setting should be selected when you need to measure
only the AC component of a signal. For example, select this setting when you are making rms or peak-to-peak voltage
measurement.
References
Press the Reference softkey to set the input references for generator.
–Voltage
Set the reference voltage value.
Ratio
Set the reference ratio value.
Frequency
Set the reference frequency value.
–Sound Lvl
Set the sound level value.
–Calibrator Lvl
Set the calibration level value value.
–Volt/FS
Set the volts/FS reference value.
Table 7-9 TSA > Project > Test > IO Configuration > Input Configuration menu description (continued)(Bluetooth)
Menu Description