— CD* Calibration data
— 000 CALIb nEW new calibration, VIEW display data, Mod
modify calibration
— 001 MAX.FSd Decimal point position, Max (scale range) and
weight unit
— 002 StEP Scale interval
— 003 dEAdLo Set dead load with weight (LoAd) or mV/V
(MVoLt)
— 004 SPAn Calibrate range with weight (LoAd) or mV/V
(MVoLt)
— 005 0-rnG Display the used zero-setting range
— 006 uVolt.d Display µV/d
— 007 WEIGHt Display current gross weight; 10-fold resolution
is possible using key.
— CP* Calibration parameter
— 010 MEAtIM Measurement time
— 011 FILtEr Digital filter
— 012 FCut Filter frequency
— 013 tStMod Test mode
— 014 WAM W&M, legal-for-trade
— 015 StStIM Standstill time
— 016 StSrnG Standstill range
— 017 TArtIM Timeout for taring/zero-setting
— 018 ZESrnG Zeroset range
— 019 ZEtrnG Zerotrack range
7.3 Switching off the device
The device is switched off/disconnected from power supply by pulling the plug.
7.4 Warm-up time
The device requires a warm-up time of 30 minutes before calibration.
7.5 Configuring and calibrating using the front-panel keys
The device can be configured directly using the front-panel keys, with the exception of
the connected xBPI weighing point.
Note:
If an xBPI weighing point is selected, "Err86" appears on the display and it can only be
calibrated via VNC.
7.5.1 Parameter table (SEtuP)
12. Use
/ and / to make changes accordingly.
13. Press "OK."
dt 086 appears on the display.
14. Press "Exit" to exit the menu.