16.2 Error messages for xBPI scales
Display VNC text Error and possible cause
Error 3 Ext. meas. device error No weight values can be read from scale:
- Scale error
Error 4 Value exceeds display The weight value is not displayed:
- Too many digits have been set.
Error 9 No values from scale No communication with xBPI scale:
- Cable break.
- Internal scale error.
- The scale is not connected to the power supply.
Error 11 No weight data No weight display:
- Another weighing point was selected.
Press
to assign the new weighing point to the device.
Requirements:
- [Serial ports parameter] - [xBPI port] - [Slot 1/2 RS
485] is selected.
Error 12 Scale not ready The scale is not ready for weighing:
- The device is in the warm-up phase.
- The device is in automatic taring mode.
- The device was switched on with the scale loaded.
Switch the device off and on again.
Error 15 Wrong serial number Serial number of scale does not match the number set in the
device.
16.3 Error messages for Pendeo load cells
Display VNC text Error and possible cause
Error 2 Overload The measuring signal is higher than Max + (x d):
- Wrong setting.
- Too much weight on the scale.
Error 4 Value exceeds display The weight value is not displayed:
- Too many digits have been set.
Error 8 Incompatible units Incompatible units:
- Incorrect calibration values e.g. due to incorrect calibra-
tion
16 Error messages X3 Process Indicator PR 5410