7.14.8 Calibrating with mV/V
The scale can be calibrated without weights. During input of the load cell mV/V value, the
acceleration of gravity at the place of installation can be taken into account.
The PR- load cell data is based on the acceleration of gravity in Hamburg, Germany:
9.81379 m/s
2
.
Note:
Error messages, calibration see Chapter
16.4.
1. Calculating SPAN value for Max and, if necessary, for the dead see Chapter 7.14.8.1.
2. Press the [by mV/V] softkey.
3. Entering the SPAN value for Max and, if necessary, for the subsequent correction of
dead load (see Chapter 7.14.10).
4. Confirm the entries.
Weight value, weight unit and measuring signal in mV/V corresponding to this
value are displayed in the line [Calibrated at].
The verification is displayed by "Setting Span mV/V…".
7 Getting started X3 Process Indicator PR 5410
Minebea Intec EN-173