[Zerotrack step]
If a weight change exceeds the adjusted value, automatic tracking does not function any
more.
Setting range for automatic tracking increments: 0.25…10 d
In "legal-for-trade" mode a value of ≤0.5 d has to be entered.
[Zerotrack time]
Time interval for automatic zero tracking.
Setting range: 0.1…25 s
At 0.0 s the zero tracking is switched off.
For use in legal metrology a value of 1 s must be entered.
[Overload]
Weighing range above the maximum load (Max) without error message.
Setting range: 0…9999999 d
For use in legal metrology a value of max. 9 d = e must be set.
[Minimum weight]
Minimum weight at which a print command can be triggered.
Setting range: 0…9999999 d
For use in legal metrology a value of at least 20 d must be set.
[Range mode]
Selection: <Single range>, Multiple range, Multi-interval
For scale range selection, see Chapter 7.14.15.2 and 7.14.15.3.
Press
to exit the menu and to save the settings.
7.16.13 Subsequent dead load correction
If the hopper/platform weight changes by an amount that is higher than the zero-setting
range; e.g.due to dead load reduction, dead load increase, or mechanical changes, the
functions for automatic zero tracking and manual zero setting no longer work.
To view the range which is already utilized by zero tracking or zero setting, in [Calibration]
press the key; this also activates 10-fold increased resolution of the weight value.
Press
again to return to the previous state.
7 Getting started X3 Process Indicator PR 5410
Minebea Intec EN-213