EasyManua.ls Logo

Minebea Intec X3 - Calibrate load cells incl. Connexx module

Minebea Intec X3
362 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
[Ambient conditions]
This parameter is used to define the ambient conditions of the scale.
Possible Selections: very stable condition, stable condition, unstable condition, very
unstable condition
[W&M]
See Chapter 7.14.15.1.
If [OIML] was selected, the device needs to warm up for 30 seconds.
[Unbal. Check deviat.]
The plausibility check is activated when the average deviation is >0%. The average
deviation of the individual load cells is calculated.
Setting range: 0…100%.
[Standstill time]
The parameters [Standstill time] and [Standstill range]are used to define the stability of
the scale (stable balance).
The input for the parameter [Standstill time] is expressed in seconds.
Valid range: 0.00…2 s
If the tome is set to "0" there is no check. The standstill time must not be less than the
measuring time.
[Standstill range]
As long as the weight fluctuations remain within this range, the device is determined to
be stable.
The input for the parameter [Standstill range] is expressed in "d."
Valid range: 0.01…10.00 dc.
[Tare timeout]
Timeout for a tare/zeroset command that cannot be executed (e.g. due to mechanical
instability of the scale, incorrect filter setting, resolution too high, standstill condition too
strict).
The input is done as seconds.
Valid range: 0.0…<2.5>…25 s.
At 0.0 s taring is only carried out when the scale is already stable.
[Zeroset range]
Define a ±range around the zero point determined by the dead load during calibration;
within this range
- the displayed gross weight can be set to zero by pressing the zero-setting key (or by
a corresponding external command), and
- automatic zero tracking is active.
Setting range: 0.00…10000.00 d
For use in legal metrology a value ≤2% of the Max must be entered, example: 60 d for
3000 e of class III.
[Zerotrack indic. range]
Indication range within which automatic zero tracking compensates deviations.
Setting range: 0.25…10000.00 d
In "legal-for-trade" mode a value of <0.5 d has to be entered.
X3 Process Indicator PR 5410 7 Getting started
EN-212 Minebea Intec

Table of Contents