Procedure:
The weight unit is taken from [Max]. The number of digits behind the decimal point is
also automatically determined when [Max] is entered.
Note:
Error messages, calibration see Chapter 16.4.
7.14.6 Determining the dead load
Note:
If a linearization was carried out (see Chapter
7.14.11), the following note appears on the
display after the [Dead load at] line is selected:
Cannot be changed here while linearization is active.
Changes cannot be made while linearization is switched on.
Only deleting of the linearization points deactivates the linearization mode!
1. Select [Scale interval] and confirm by pressing or .
A selection window opens.
2. Select the scale interval (1 d) and confirm with .
The verification is displayed by "Setting Scale interval…".
The scale interval (d) is then calculated, based on the Max weight value.
X3 Process Indicator PR 5410 7 Getting started
EN-170 Minebea Intec