7.17.3 Calibration procedure
During calibration, no data is changed in the Connexx module. The calibration data and
parameters are saved in the device. The unique serial numbers of the connected load cells
are monitored.
For the calibration the following order must be followed:
- Search for load cells, see Chapter 7.17.4.
- Assign load cells, see Chapter 7.17.5.
- Recalibrate: Maximum capacity with weight unit, scale interval, dead load, calibration
weight, see Chapter 7.17.6.
- Perform a corner correction if necessary; see Chapter 7.16.10.3.
Note:
For further information about calibrating weighing points, see Chapter
7.14.3.
7.17.4 Search for load cells
1. Select -[Weighing point] - [Weighing point A].
2. Press the [Assign] softkey.
The following window opens:
3. Press the [Search] softkey.
7 Getting started X3 Process Indicator PR 5410
Minebea Intec EN-215