Press the [Cancel] softkey to accept and display the existing values.
A prompt window opens.
4. Press the [Continue] softkey to start a new search process.
[Type]
Type of load cells
[No of LC]
Number of load cells
[LC 1…n]
Serial number/name of the load cells
[WP serial number]
Weighing point serial number(is displayed after search)
A window with load cell information opens
5. Press the [View] softkey.
The load cells are displayed with their item number, serial number, and load.
6. Select the desired load cell and press the [Info] softkey.
Note:
If load cell names have been assigned (see Chapter 7.17.7), the view can be
switched with the [by name] softkey.
The load cell data is displayed.
7. Press
to exit the menu and save.
X3 Process Indicator PR 5410 7 Getting started
EN-216 Minebea Intec