Data type Description Value range
WSTRING variable-long wide cha-
racter string
max. 255 characters (Unicode)
BYTE bit-sequence 8
WORD bit-sequence 16
DWORD bit-sequence 32
LWORD bit-sequence 64
13.3 Addressing
The SPM table can be addressed via different counts. Bit addressing is used to count the
individual bits (MX). Byte addressing is used to count individual bytes (MB), whereby, e.g.
bits MX0…MX7 are identical to byte MB0.
Code Data type Address example
%ML LWORD L21
%MD DINT D42…43
%MW WORD W84…87
%MB BYTE B168…175
%MX BOOL (bit) X1344…1407
13.4 System data
SPM address Data type R/W Function
X0…X2 BOOL R Digital input 1…3
X8…10 BOOL R Digital output 1…3
X16…18 BOOL R Standard only: Output limit 1…3
B4
X32
X33
X34
X35
X36
X37
X38
X39
BYTE
BOOL
BOOL
BOOL
BOOL
BOOL
BOOL
BOOL
BOOL
R
R
R
R
R
R
R
R
R
Indicator status
ADC error
> Max (FSD = Full Scale Deflection)
> Max + permitted range (OVL)
< zero
Zero ±¼ d
Within the zeroset range (ZSR)
The weight is stable
Weight < zero or > Max (FSD = Full Scale Deflection)
X3 Process Indicator PR 5410 13 SPM
EN-336 Minebea Intec