10 ModBus protocol
10.1 General description
The ModBus protocol implemented in the device enables rapid, simple, and reliable
communication between a PC or SPS and up to a maximum of 127 devices.
The ModBus protocol allows access to all data published in the SPM table of the relevant
application.
Implementation:
The functions 1, 2, 3, 4, 5, 6, 8, 15, and 16 are supported.
Bits can only be read or set individually or in groups of eight.
10.2 SPM data in PR 1612 ModBus mode
For access via ModBus, the PR 1612 ModBus mode is switched on via - [Serial ports]-
[ModBus-RTU]- [Param]- [ModBus mode]- [PR 1612 ModBus].
Read data
Byte address Weight in 32-bit integer format
60…63
64…67
68…71
Gross weight
Net weight (gross if not tared)
Tare weight (0 if not tared)
Read data
Address Read "Word" (display in binary format)
W201 0E00000T 00000000
E: ADC error
T: Instrument is tared.
W203 000S00MZ 00000000
S: Device is in setup mode.
M: Weight is stable (standstill).
Z: Weight within ¼d of 0
W204 0000D000 00000000
D: Gross weight <0 or >Max (dimmed)
W205
TA000000 00000000
T: Instrument is tared.
A: Analog test is active.
W231 00021MRZ 00000000
2: Limit 2
1: Limit 1
M: Weight is stable (standstill).
R: Weight within zero-setting range
Z: Weight within ¼d of 0
10 ModBus protocol X3 Process Indicator PR 5410