Saving and exiting
7.5.3 Changing the dead load of the internal weighing point using the front-panel keys
Note:
If linearization is active (see Chapter
7.14.11), Error 91 appears on the display.
If modifying the mechanics results in the dead load (weight of the empty scale/container)
being changed, proceed as described in the following.
Start
Changing the dead load (Cd 003)
38. Press "Exit."
SAVE appears on the display.
39. Press OK.
YES appears on the display.
40. Press OK.
SAVE ... appears on the display during the save process.
41. Press "Exit" to exit the Setup menu.
42. If calibration parameters CP 010–CP 026 are not intended to be set or are already set,
the CAL switch must be closed now; see Chapter 7.1.3.1.
1. Switch on the device; see Chapter 7.2. Allow for the warm-up phase for the device to
be completed.
2. Open the CAL switch; see Chapter 7.1.3.1.
3. Press .
SEtuP appears on the display.
4. Press OK.
The "Cd" calibration menu flashes.
Cd 000 appears on the display.
5. Press to change the parameter number ("000" flashes).
6. Press to select the parameter number.
7. Press OK.
dEAdLo appears on the display.
8. Press OK.
7 Getting started X3 Process Indicator PR 5410
Minebea Intec EN-133