Continue for "dead load with empty scale":
Defining the span (Cd004)
Continue for "range with weights":
MVolt appears on the display.
24. Press OK.
unLoAd appears on the display.
25. Empty the scale.
26. Press OK.
A weight value for the dead load appears on the display.
27. Press OK.
00000.0 kg appears on the display.
The dead load is set.
28. Press OK.
Cd 004 appears on the display.
29. Press OK.
SPAn appears on the display.
30. Press OK.
LoAd (range with weights) appears on the display.
31. Press to select the "mV/V" entry if necessary.
MVolt appears on the display.
32. Press OK.
33. Place the calibration weight on the scale.
A weight value that has not been calibrated yet appears on the display.
34. Press OK.
WEIGHt appears on the display.
35. Press OK.
The max. calibration weight of 02000.0 g appears on the display.
36. Use the cursor keys to set the calibration weight value.
37. Press OK.
Cd 005 appears on the display.
X3 Process Indicator PR 5410 7 Getting started
EN-132 Minebea Intec