Display VNC text Error and possible cause
Error 9 No values from scale No communication with Pendeo load cell:
- Cable break.
- Internal scale error.
- The junction box is not connected to the supply voltage.
Error 12 Scale not ready The scale is not ready for weighing:
- At least 1 load cell gives an error status or is defective
(no communication).
Error 13 Wrong configuration The number of load cells does not match the configuration.
Error 15 Wrong serial number Serial number of scale does not match the number set in the
device.
16.4 Error messages during calibration
Message Possible causes
This message displays, if the maximum load is too high.
The maximum load of the scale can be increased retroactively. When the
capacity is reduced, however, a message is displayed when the new ma-
ximum load is lower than the CAL weight ([Calibrated at]).
This message is displayed if the selected resolution is too low, e.g., 5 kg.
This message is displayed if the selected resolution is so high that less
than 0.8 internal counts per scale interval (d) and/or 0.5 µV/e are avai-
lable if legal-for-trade acc. to OIML/NSC.