Message Possible causes
This message displays if the maximum load is not an integer multiple of
the scale interval.
Weight units don't match, e.g., subsequent change for [Max] from kg in-
to lb.
This message is displayed when the dead load entered in mV/V plus
scale interval in mV/V is higher than 3 mV/V (= 36 mV).
This message displays, if the scale is not stable.
Remedy
- Check the mechanical function of the scale.
- Adapt the filter setting; reduce the resolution.
- Adapt the standstill conditions.
This message is displayed when the measurement signal is negative
when determining the dead load with [by load].
Cause
Load cell connected with wrong polarity, or defective.
This message is displayed when the dead load entered is ›5 mV/V.
This message displays, if the scale is not stable.
Remedy
- Check the mechanical function of the scale.
- Adapt the filter setting; reduce the resolution.
- Adapt the standstill conditions.
16 Error messages X3 Process Indicator PR 5410