Display Error and Possible Cause
Err 90 No printer interface assigned.
Err 91 Linearization is active.
Err 92 xBPI interface cannot be opened.
Err 93 Too many Pendeo load cells were found.
Err 94 Search for Pendeo load cells is defective.
Err 95 Communication error with Pendeo load cells.
Err 96 Not all Pendeo load cells are the same type.
Err 97 Dead load cannot be set for Pendeo load cells.
Err 98 Weighing point cannot be assigned to the Pendeo load cell.
Err 99 xBPI communication is not assigned.
16.6 Error numbers @ "LAST_ERROR"
16.6.1 Weighing point error
Number Display Cause
2 in use Weighing point is in use, e.g., by an application.
6 test active Test is active, no weights.
7 cali active Calibration is active, no weights.
8 no standstill No standstill of the scale.
13 tare is active Tare is active.
15 weight is dimmed Weight is not legal-for-trade (<0 or >max.).
16 weight has error Weight error.
17 scale not ready Scale is not ready.
18 cannot tare below zero Taring below zero is not possible.
102 timeout Only on xBPI scales: Time limit exceeded when sending a com-
mand to the scale.
142 calibration active During calibration, taring and zeroing is not possible.
147 no zeroset Zeroset outside of the zero setting range is not possible.
149 Busy The scale is currently busy with another query.
255 hardware error Weighing point is faulty.
16.6.2 Error in the "EasyFill" application
Number Display Cause
0 No error.
1 fatal error Weight error; weighing point is faulty.
X3 Process Indicator PR 5410 16 Error messages
EN-352 Minebea Intec