7.14.9 Calibrating with load cell data (smart calibration)
If the scale is not used in legal metrology, calibration without weights can be performed.
The easiest method is the one using load cell data without calculation.
[Number of load cells]
Number of load cells connected in parallel
Input: 1, 2…<4>…9, 10
[max. capacity of load cell]
Maximum capacity E
max
of a load cell (not the total maximum capacity of the scale!)
Input: For the value refer to the technical data of the load cell.
[Gravity]
Gravity at place of installation
Default is the value for Hamburg, Germany: 9.81379 m/s
2
.
[Hysteresis error]
Hysteresis error
When switching from [not specified] to [specified] values for [Correction A/B] must
be entered. For this data refer to the load cell certificate.
1. Press the [by data] softkey.
7 Getting started X3 Process Indicator PR 5410